INTRODUCTION
An indicator is a weak acid or base whose conjugate forms have different
colors. In this experiment, the indicator, also used as a biological stain,
is Neutral Red (3-amino-7-dimethylamino-2-methylphenazine hydrochloride)
(NR), which has the structure (protonated form) and acid dissociation equilibrium
shown below.
HNR+ + H2O = H3O+
+ NR
Figure 1. The structure and acid dissociation of Neutral Red
(3-amino-7-dimethylamino-2-methylphenazine hydrochloride).
The thermodynamic acid dissociation constant for HNR+ is given in terms of activities
which reduces to
because the pure solvent is taken to be standard state (a = 1). The activity is a concept of "effective concentration" which is envoked to correct for non-ideal solution behavior which is expected in aqueous electrolytes. This non-ideality may be represented by an activity coefficient, g , which is the ratio of the activity over the physical concentration.
where ci is the molar concentration of species i. The equation for Ka is, therefore,
The activity coefficient of a neutral molecule in a dilute solution may be taken as unity. The activity coefficient of an ion is determined by the total ionic atmosphere of the solution, called the ionic strength, µ,
where ci is the concentration of ion i, zi is its charge, and the sum is taken over all ions in solution. Several equations can be used for calculation of activity coefficients, but a very common one is the extended Debye-Huckel equation:
where z is the charge of the ion of interest, and a is the radius of the solvated ion (ion plus its tightly bound sheath of water molecules; a = ca. 900 pm for H3O+; a = ca. 700 pm for a large cation like HNR+). A and B are factors which depend on the density (A only) and dielectric constant of the solvent and the temperature. Values of at 25 oC for various ionic strengths and a's are conveniently obtained from the literature [1].
If the ionic strength is held constant for all measurements involving Kaµ, a "concentration" dissociation constant, valid only at that ionic strength, may be used:
In this experiment, the Kaµ=0.10 will be measured at a total ionic
strength of 0.10M, maintained by adding an inert salt, NaCl, to all solutions.
The [H3O+] will be controlled by a buffer of
NH3OH+/NH2OH created by incremental addition
of standard 0.10M NaOH to a known number
of moles of either NH3OH+
(Kaµ=0.10 for NH3OH+ at 25oC
and µ = 0.1M is 1.07 x 10-6). As described further below, the addition of NaOH will increase
the solution volume, necessitating a correction for dilution in the calculation of all concentrations.
The [NR]/[HNR+] ratio will be determined spectrophotometrically. If a solution with a total indicator concentration of cT is made very acidic, all of the indicator exists as HNR+. The absorbance of the solution at a given wavelenght of light, l, is given by
where eHNR+ is the molar absorptivity of HNR+ at wavelength l and b is the cell path length. If, by making the solution very basic, the same concentration of indicator is converted entirely to the NR form, the absorbance at the same wavelength is given by
where eNR is the molar absorptivity of NR. At an intermediate pH the absorbance is
where the total concentration can be defined under any conditions as:
For a given cT, equations 8-11 can be combined to give
If possible, the ratio should be evaluated at multiple wavelengths, including one where HNR+ absorbs appreciably but NR does not, and one where NR is a much stronger absorber than HNR+, and one where the two species absorb at approximately the same extinction (What is this point called??). The pCH's (-log[H3O+]) of the solutions should be in the transition range of the indicator, so that both HNR+ and NR exist in appreciable concentration. With a spectrophotometer having a conventional sample holder (cuvette) a series of separate solutions must be prepared. However, if the instrument is equipped with a fiberoptic probe, the experimental procedure can be greatly simplified. A single solution containing the indicator plus NH3OH+ is prepared and known increments of NaOH are added to convert the respective acid to its conjugate base form. This addition changes the total volume and decreases cT, but equation 12 can still be used if the absorbance values are corrected for the dilution effect.
Referring to equation 7, Kaµ is numerically equal to [H3O+] when [NR]/[HNR+] is one. This can be evaluated graphically by converting equation 7 to logarithmic form:
Thus, a plot of the logarithm of the [NR]/[HNR+] ratio versus pCH (-log [H30+]) has an ideal slope of unity. If this condition holds, the y-intercept gives the negative of pKa directly. However, usually the experimental slope is not exactly 1.000. Extrapolation from the pCH's of the measurements to pCH = zero (which is typically a factor of 106 in concentration for this study) results in a very large error in pKaµ. On the other hand, data are obtained on both sides of the x-intercept, which is determined with very little uncertainty. Therefore, better results are obtained by solving the least-squares equation for the x-intercept (the value of the pCH when [NR]/[HNR+] is one).
DESCRIPTION OF THE EXPERIMENT
The pKa of Neutral Red will be determined in a strictly
aqueous environment, by measuring the equilibrium dissociation spectrophotometrically
as a function of pH at fixed ionic strength.
PROCEDURE
(Note: This procedure has been designed for use on the Spectral InstrumentsTM
model SI 440 spectrophotometer, which is equipped with a fiberoptic probe
and a CCD detector for convenient and rapid operation. The fiberoptic
probes must be handled carefully to avoid damage.)
1. Turn on the system at the power strip, press the ON pad on the temperature
controller, and flip the toggle switch on the spectrophotometer module
to the ON position. Click 'Cancel' when the password request appears. Click
the 'SI 430-440', Windows icon. The SI main front panel display will appear.
Click 'Initialize' (lower right). Within a few seconds, an audible internal
switch will be heard. Then, click 'Wavelength Range' and make sure it is
set to 400 to 720 nm with the lamp cutoff wavelength at 400 nm. Make sure
that the 'Precision' is set to 'Normal' (20 readings of the CCD are signal
averaged). Allow time for warmup. (The tungsten lamp requires 30 minutes
to stabilize; baseline drift may occur otherwise.) Insert a diskette in
the A drive.
2. To choose wavelengths and evaluate AHNR+ and ANR, it is necessary to prepare two indicator solutions--one containing only HNR+ and the other containing only NR--for each solution system. Prepare solutions in 50 mL volumetric flasks as follows, using 0.10 M NaCl for all final dilutions:
HNR+: 5 mL Neutral Red (provided as a 0.01% solution in 0.10 M NaCl) + 5 mL 0.10 M HCl
NR: 5 mL NR + 5 mL 0.10 M NaOH
Transfer some of each solution to appropriately labelled beakers. (These
solutions do not require temperature control.)
3. Immerse the fiberoptic probe in deionized H2O and stir
gently with the probe to dislodge any bubbles. Click 'Blank' to
acquire the blank spectrum to be used by the software for absorbance calculation.
These data are used until the wavelength range is changed or until a new
blank is acquired. Click 'Lock Blank' to prevent accidental repeat of the
blank measurement. (If a new blank is needed later, first click 'Unlock
Blank'.)
4. Remove the probe from the water and gently blot the probe
with a bed of KimwipesTM to remove liquid remaining on the mirror.
Gently dry the outside of the probe.
5. Acquire and save the spectrum for each solution as follows: Immerse
the dry probe in the sample, and click 'Sample'. Then click the red 'Save'
box on the lower right of the display, and save the spectrum on your floppy
(A drive) using an appropriate filename. (The 'Save' box will return to
background color after the spectrum has been stored.).
6. After you have obtained the two spectra for each system, click the
'Open' box (next to 'Save') and recall the HNR+ spectrum. Under
'Reference Spectrum' click 'Set' and toggle the adjacent reference button
to 'On'. Now recall the NR spectrum for the same system. Use the cursor
to locate the lmax for each form
and the absorbance of each form
at each wavelength (two total wavelengths and four total absorbance values).
Choose several other wavelengths at
which to make absorbance measurements, including a wavelength at which minimal
absorption occurs in either species occurs which can be used for baseline correction.
(The baseline tends
to shift a little during the experiment. To correct for this, the absorbance
at 700 nm, for example, where neither form absorbs, can be subtracted from each measured
absorbance to improve the accuracy of the analysis.) Click 'Print' on the right panel (not
on the Windows File pulldown) to obtain a hard copy. Toggle the reference
button to 'Off'.
7. Obtain a 100-1000 µL Eppendorf pipettor, and practice using
it on the 200 and 1000 µL settings. Determine the exact volumes dispensed
by weighing samples of water. If necessary, adjust the micrometer dial
to dispense volumes of 0.2000±0.0001 mL and 1.000±0.001 mL.
Use the same tip for all calibrations.
8. Solutions for Kaµ determination will be prepared directly
in the jacketed vessel. pipet 5 mL NR, 5 mL 0.1 M NH3OH+Cl-
(record the exact concentration), plus 40 mL (2x20mL) 0.10 M NaCl (50 mL
total). Immerse the probe and turn on the stirrer. Locate the 'Wavelength'/
'Absorbance' table on the computer display, and enter the two chosen wavelengths
plus 700 nm. Then click 'Sample'. Record the three absorbances from the
table, and store the spectrum on the floppy. Click 'Save Setup' to keep
the same wavelengths in the table for all spectra.
9. Add 2000 µL (2.000 mL) of the standard base (record the exact
concentration). Obtain and save the spectrum and record the three absorbances.
10. Repeat this procedure using increments which will give the total
volumes listed in Table 1. Save each new spectrum to your floppy disk.
Use the same pipet tip for all additions of
NaOH. Rinse and blot dry the probe and the cell.
11. When
you have finished with all the additions, click on the "Export Files"
button located to the right of the Save button. A window will open called
"File Dialog". Select the A: drive and select the folder containing
the saved spectra. A list of all your saved files should appear.
Using the shift key and the left-mouse button, select all the spectra
files and click "OK." A new window will open which will give you the
option of saving the files in either a text or excel spreadsheet format.
Choose the one most applicable to you.
12. When you have completed all measurements and created your spreadsheet file(s),
ask your TA if the spectrometer is to be shut down. If so, click 'File' and 'Close'
in the upper left corner of the display. When the Windows descktop appears,
click 'Start' and 'Shut Down'. Answer 'Yes' to the query. Turn off the
spectrophotometer module, the water bath, and the power strip.
Table 1. Suggested total volumes (in mL) of standard 0.1 M
NaOH.
| 25oC, µ = 0.10 |
| 2.0, 2.2, 2.4, 2.6, 2.8, 3.0, 3.2, 3.4, 3.6, 3.8, 4.0, 4.2, 4.4, 4.6, 4.8 |
DATA ANALYSIS
1. Derive equation 12 in the Introduction section of your report.
2. Correct all absorbance readings for the baseline offset at 700 nm. Then correct for dilution by multiplying by the ratio of the total volume to the initial volume.
3. Use equation 12 to calculate the [NR]/[HNR+] ratio resulting from each addition of standardized NaOH. (Omit the data for 0.000 mL.) For Systems 2 and 3, use the data for both wavelengths. Omit the lmax for NR. (In water, NR tends to form aggregates (7), leading to a complication in the analysis of the data due to the change in optical properties of the dye upon aggregation. Why does HNR+ not aggregate?)
4. Calculate the concentrations of buffer species (NH3OH+ & NH2OH. Calculate [H3O+] from the buffer properties.
4. Determine the pKaµ for Neutral Red in each system and estimate the error limits in the determination by a graphical method (equation 13).
5. Consult reference 1 or another suitable source for the activity coefficents
of H3O+ and HNR+, and calculate the thermodynamic
Ka for this indicator and estimate the error limits of this measurement and the correction for
non-ideal activities in the system.
DISCUSSION AND CONCLUSION
What is an indicator and how do they work? How do the values of the pKapKa's of indicators determine
their use? How do uncertainties in the pKa values of an indicator translate into
uncertainties in experiments that use them? How important are corrections for activities in titrations?
What theories exist for the estimation of activity coefficients in electrolytes? How well do they work?
REFERENCES
1. Harris, D.C. Quantitative Chemical Analysis, 5th Ed., W.H.
Freeman: New York, 1999, p 180.
2. van Stam, J.; Depaemelaere, S.; De Schryver, F. J. Chem. Educ.,
1998, 75, 93-98.
3. Hartley, G.S.; Roe, J.W. Trans. Faraday Soc., 1940, 36,
101-109.
4. Mukerjee, P.; Banerjee, K. J. Phys. Chem. 1964, 68,
3567-3573.
5. Fernández, M.S.; Fromherz, P. J. Phys. Chem. 1977,
81, 1755-1761.
6. Hartland, G.V.; Grieser, F.; White, L.R. J. Chem. Soc., Faraday
Trans. 1, 1987, 83, 591-613.
7. Drummond, C.J.; Grieser, F.; Healy, T.W. J. Chem. Soc., Faraday
Trans. 1, 1989, 85, 551-560.